Before Calibration Adjustment; Maintenance Mode; Password Entry; Change The Password - Fluke CNX a3000 Calibration Manual

Wireless ac clamp
Hide thumbs Also See for CNX a3000:
Table of Contents

Advertisement

Before Calibration Adjustment

Before the Product calibration can be adjusted, you must go through the maintenance
mode menu and enter your password.

Maintenance Mode

The Product maintenance mode can be used to set different parameters on the Product
that include auto power off, backlight adjustment, and calibration. To use the
maintenance mode:
1. With the Product OFF, push and hold .
2. Push . Keep  pushed until all the display segments are shown.
3. Release  and .
The Product is now in maintenance mode.

Password Entry

To go to the calibration mode, push  until CAL is shown. You will need to enter a
password to access calibration mode.
To enter the password:
1. Push  and the CAL version is shown. For example n002.
2. Push  to show "????". The first "?" flashes.
3. Push  to change the flashing "?" to the first digit of your password
(default: 1234).
4. Push  to confirm your choice. The subsequent "?" flashes.
5. Do steps 3 and 4 again to enter the subsequent digits of the four-number password.
6. When all of the correct digits are entered, push  to confirm the input.
If the correct password is entered, "C-01" is shown. If the incorrect password is
entered, "????" is shown and the password must be correctly entered to go to the first
calibration point, "C-01".

Change the Password

If you change the password and then lose it, see the "Restore the Default
Password" section.
To change the password:
1. Do steps 1 through 5 in the "Password Entry" section.
2. Before you push  to confirm your final input (step 6), push  to show "----" on
the display. The first "-" flashes.
3. Push  to change the first "-" to the first digit of your new password.
4. Push  to confirm your choice. The next "-" flashes.
5. Repeat steps 3 and 4 to enter the subsequent digits of the new four-number password.
6. When the correct digits are entered, push  to confirm the input and change the
password. If the Product has been calibrated, it will go to normal measurement
mode, or it will show "donE".
riešenia na presné meranie
Note
http://www.elso.sk
Wireless AC Clamp
Before Calibration Adjustment
9

Advertisement

Table of Contents
loading

Table of Contents