Page 2
LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling.
Table of Contents Title Page Introduction ........................1 How to Contact Fluke ..................... 1 Standard Equipment....................... 2 Safety Information ......................2 Hazard Location Information/Approvals ..............3 Special Conditions for Safe Use ................3 Symbols ........................4 Display and Buttons ....................... 5 Operation ........................
Page 4
700G Series Users Manual Set Damping ......................7 Set Sample Rate ....................7 Set TARE ......................7 Function Lock ....................... 8 Supervisory Mode ..................... 8 Available Pressure Ranges ..................8 How to Set a Custom Engineering Unit or Scale............9 Battery Life ........................
Introduction The 700G Series Pressure Gauges (the Product) are high-accuracy digital pressure test gauges. Accurate to 0.05 % FS, the Products can be used as a calibration reference, or in any application where high-accuracy pressure measurement is required. The Product features user-configurable functions that include: •...
700G Series Users Manual • Standard Equipment Read all safety Information before you use the Product. The Product ships with: • Do not use the Product in damp or wet • Protective Cover environments. • Three AA Batteries (installed) WCaution •...
These areas are also any damage that can cause a safety concern. If referred to as hazardous locations, see NFPA 70 Article necessary, return the Product for evaluation to Fluke. 500. Refer to the How to Contact Fluke section.
Conforms to relevant North American Hazardous voltage. Risk of electrical shock. Safety Standards. Do not dispose of this product as Pressure unsorted municipal waste. Go to Fluke’s website for recycling information. Conforms to relevant Australian standards. Conforms to ATEX requirements Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176...
Pressure Gauge Display and Buttons Display and Buttons Table 2. Display and Buttons The Display and Buttons are shown in Figure 1. The Item Function Buttons are explained in Table 2. Push to turn the Product on. Push again to ...
700G Series Users Manual Operation Table 2. Display and Buttons (Cont) The subsequent sections tell you how to operate the Item Function Product. Push to turn on the Product. The analog bar graph at the bottom of the display shows Push to go to setup and configuration ...
Pressure Gauge Operation Set Auto Off Set Sample Rate Auto Off can be set in 1-minute increments from 1 to 30 This function finds how often pressure is sampled and the minutes or you can turn off the function for continuous display is updated.
700G Series Users Manual Function Lock each parameter. To move to the next parameter, push . When set, access to each of the settable parameters above can be turned “off” to prevent unauthorized You can access, lock, or unlock these functions: configuration changes.
Pressure Gauge Battery Life How to Set a Custom Engineering Unit or Scale Maintenance The last menu selection in Supervisory mode is How to Clean the Product SET FACTR. You can set a multiplier factor from 0.001 to Clean the Product with a soft cloth dampened with water 100 to make a custom scale.
700G Series Users Manual How to Change the Batteries XWWarning To prevent possible electrical shock, fire, or personal injury, batteries must only be changed in an area known to be non- hazardous. Explosion hazard. To change the batteries, see Figure 2: Use a Phillips screwdriver to loosen the captive screw on the battery door.
Pressure Gauge Accessories Accessories RS-232 Interface The Product includes an RS-232 interface. Remove the Product holster and the input jack is on the back of the Product. You can use serial communication to configure and calibrate the Product and move measurement data from the Product to a PC.
700G Series Users Manual Specifications Available Input Ranges See PI Ranges and Resolution for available ranges in psi plus equivalent ranges and resolution for all engineering units. Accuracy Positive Pressure ................±0.05 % FS Vacuum ..................±0.1 % FS Temperature Compensation ............15 °C to 35 °C (59 °F to 95 °F) to rated accuracy Note: For temperatures from -10 °C to 15 °C and 35 °C to 55 °C, add .003 % FS/°C Media Compatibility 15, 30 psi ..................
Pressure Gauge Specifications Mechanical Specifications Dimensions ..................11.4 x 12.7 (cm), depth = 3.7 cm (4.5 x 5 (in), depth= 1.5 in) Pressure Connection ................¼ in NPT Male Housing .................. Cast ZNAL Display 5-1/2 Digits, 16.53 mm (0.65 in) high 20-Segment bar graph, 0 to 100 % Power Battery ...................