Electromagnetic immunity

This DR Detector is intended for operation in the electromagnetic
environment given below. The user of the DR Detector should ensure that it is
used in such an environment.
Resistance to Jam-
ming Test
Discharge of static
electricity in accord-
ance with
IEC 61000-4-2
Fast transient elec-
trical disturbance
variables / bursts in
accordance with
IEC 61000-4-4
Impulse voltages
(surges) in accord-
ance with IEC
61000-4-5
Voltage break-
throughs, short
term interruptions
and variations in
the voltage supplied
in accordance with
IEC 61000-4-11
XD 10, XD⁺10 | Remarks for HF-emission and immunity | 131
IEC 60601
Level of
Test Level
Agreement
± 8 kV contact
± 8 contact
discharge
discharge
± 15 kV air
± 15 kV air
discharge
discharge
± 2 kV for net-
± 2 kV for net-
work leads
work leads
± 1 kV for en-
± 1 kV for en-
try and outlet
try and outlet
leads
leads
± 1 kV push-
± 1 kV push-
pull voltage
pull voltage
± 2 kV com-
± 2 kV com-
mon mode
mon mode
voltage
voltage
100% reduc-
100% reduc-
tion for 0.5 cy-
tion for 0.5 cy-
cle at 0, 45, 90,
cle at 0, 45, 90,
135, 180, 225,
135, 180, 225,
270 and 315
270 and 315
degrees
degrees
100% reduc-
100% reduc-
tion for 1 cycle
tion for 1 cycle
30% reduction
30% reduction
for 25/30 cy-
for 25/30 cy-
cles at 0 de-
cles at 0 de-
gree
gree
Electromagnetic
Environment
Guidelines
Floors should con-
sist of wood, con-
crete or ceramic
tiles. The relative
humidity must be at
least 30%, if the
floor is made of syn-
thetic material.
The quality of the
voltage supplied
should correspond
to a typical commer-
cial or clinical envi-
ronment.
The quality of the
voltage supplied
should correspond
to that of a typical
commercial or clini-
cal environment.
The quality of the
voltage supply
should correspond
to that of a typical
commercial or clini-
cal environment. If
the user wants the
DR Detector to work
continuously, even
when the energy
supply is interrup-
ted, it is recommen-
ded to use an ener-
gy supply free of in-
0432A EN 20210426 1355