DR detector not ready for exposure

Details
Cause
(only if the DR detec-
tor is configured in
client mode and used
on more than one
MUSICA Acquisition
Workstation)
Brief Solution
Cause
(only if the DR detec-
tor is configured in
access point mode)
Brief Solution
Cause
(only if the DR detec-
tor is configured in
access point mode)
The DR detector is turned on. The DR detector status
icon on the DR Detector Switch is not green.
The DR detector is not connected to the MUSICA Ac-
quisition Workstation.
Connect the DR detector to the MUSICA Acquisition
Workstation.
The AP mode button has been pushed accidentally.
Check the detector status display.
• If the DR detector is in client mode, the connec-
tion status is displayed.
• If the DR detector is in access point mode, the ac-
cess point icon is displayed.
If the status does not match the way it should com-
municate to the workstation, press and hold the AP
mode button for 3 seconds.
The DR detector switches to the other communica-
tion mode.
The MUSICA Acquisition Workstation is not connec-
ted to the DR detector via the wireless network.
XD 10, XD⁺10 | Problem solving | 113
0432A EN 20210426 1355