Replacing the battery

To replace the battery, a flat-head screwdriver is required.
For regular use, the other means for charging the DR detector are
recommended. Battery replacement may be required if the battery is broken
or if there is no time for regular charging and a back-up battery is available.
1. Turn off the DR detector.
Press and hold the power button (approx. 3 seconds).
2. Use the flat-head screwdriver to loosen the knobs that secure the battery
cover plate.
Turn the knobs counterclockwise a quarter turn.
3. Take off the cover plate.
4. Remove the battery.
5. Insert the fresh battery.
Make sure that the orientation of the battery is correct, so that the
electrical contacts of the battery connect to the electrical contacts of the
DR detector.
6. Put the cover plate in place.
7. Use the flat-head screwdriver to secure the battery cover plate.
8. Turn the knobs clockwise a quarter turn.
9. Turn on the detector.
Press and hold the power button for 1 second.
XD 10, XD⁺10 | Advanced Operating | 103
0432A EN 20210426 1355