Electromagnetic immunity

This DR Detector is intended for operation in the electromagnetic
environment given below. The user of the DR Detector should ensure that it is
used in such an environment.
Resistance to
Jamming Test
Discharge of
static electricity
in accordance
with
IEC 61000-4-2
Fast transient
electrical
disturbance
variables /
bursts in
accordance with
IEC 61000-4-4
Impulse voltages
(surges) in
accordance with
IEC 61000-4-5
Voltage
breakthroughs,
short term
interruptions
and variations in
the voltage
supplied in
accordance with
IEC 61000-4-11
DR 18M, DR 24M | Remarks for HF-emission and immunity | 75
IEC 60601 Test
Level of
Level
Agreement
± 6 kV contact
± 6 contact
discharge
discharge
± 8 kV air
± 8 kV air
discharge
discharge
± 2 kV for
± 2 kV for
network leads
network leads
± 1 kV for entry
± 1 kV for entry
and outlet leads
and outlet leads
± 1 kV push-
± 1 kV push-
pull voltage
pull voltage
± 2 kV common
± 2 kV common
mode voltage
mode voltage
< 5% Ur
< 5% Ur
(> 95%
(> 95%
breakthrough of
breakthrough of
Ur) for ½
Ur) for ½
period
period
40% Ur (> 60%
40% Ur (> 60%
breakthrough of
breakthrough of
Ur) for 5
Ur) for 5
periods
periods
70% Ur (30%
70% Ur (30%
breakthrough of
breakthrough of
Ur) for 25
Ur) for 25
periods
periods
Electromagnetic
Environment
Guidelines
Floors should consist
of wood, concrete or
ceramic tiles. The
relative humidity
must be at least 30%,
if the floor is made of
synthetic material.
The quality of the
voltage supplied
should correspond to
a typical commercial
or clinical
environment.
The quality of the
voltage supplied
should correspond to
that of a typical
commercial or
clinical environment.
The quality of the
voltage supply
should correspond to
that of a typical
commercial or
clinical environment.
If the user wants the
DR Detector to work
continuously, even
when the energy
supply is
interrupted, it is
recommended to use
an energy supply
0380A EN 20180703 1429