Identifying problems

Please refer to the details of following symptoms or error messages. If the
problem persists, turn off the detector and consult your sales representative or
local dealer.
Symptom
The detector will not
turn on.
The status indicator of
the System Control Unit
does not light up.
The status indicator of
the System Control Unit
does not light up in
green.
The detector status dis-
play shows an error
number (e.g. "ERR_01").
A fully charged battery is
consumed quickly.
WARNING:
Improper changes, additions, maintenance or repair of the
system can lead to personal injury, electrical shock and damage
to the equipment. Safety is only guaranteed when changes,
additions, maintenance or repairs are carried out by an Agfa
certified field service engineer. A non certified engineer
performing a modification or service intervention on a medical
device, acts on his own responsibility and makes the warranty
void.
Cause
The battery is
not attached.
The battery pack
is not charged.
The battery pack
is broken.
The power cord
is unplugged
from the AC out-
let.
A hardware error
has occurred.
An error has oc-
curred.
The battery ca-
pacity decreases.
XD 10, XD⁺10 | Problem solving | 117
Remedy
Attach the battery.
Fully charge the battery pack.
Replace the battery pack.
Connect the plug to the AC
outlet firmly. If it still does not
work, replace the System Con-
trol Unit.
Turn off the System Control
Unit and turn it on again. If it
still does not work, replace the
System Control Unit.
Turn off the DR Detector and
turn it on again. Follow the in-
structions for starting up the
DR detector.
If the error persists call Serv-
ice.
The DR Detector battery can
deteriorate because of its char-
acteristics and structure. For
0432A EN 20210426 1355