Switching temporarily to client mode

The DR detector is configured in access point mode.
The DR detector can be temporarily switched to client mode, to connect to
another MUSICA Acquisition Workstation. The detector can be configured
upon installation with a single MUSICA Acquisition Workstation (with access
point or System Control Unit) to which it will connect in this status.
To switch the DR detector to client mode:
1. Take the DR detector near the MUSICA Acquisition Workstation with the
access point.
2. Press and hold the AP mode button for 3 seconds.
Figure 29: AP mode button
The detector status display shows "STA Mode Start", indicating that the
client mode is being activated.
3. Wait until the connection to the MUSICA Acquisition Workstation is
established.
The detector status display shows the signal strength.
4. Perform the examinations.
5. To return the DR detector to access point mode, press and hold the AP
mode button for 3 seconds.
The detector status display shows "AP Mode Start", indicating that the
access point mode is being activated.
After a while, the detector status display shows the access point icon.
Related Links
Configuration
on page 14
XD 10, XD⁺10 | Advanced Operating | 109
0432A EN 20210426 1355