Service setup

9.5 Software parameters

9.5.1 Configuration file (config.mcf)

Ini-key
b_all_tools
b_aquinity_start
b_lf_sensor_1
b_lf_sensor_2
b_load_save_backup
b_tank_level_limit
b_temp_too_high
dbl_init_q1_value
dbl_init_q2_value
dbl_init_q3_value
dbl_init_q4_value
dbl_init_tank_level
dbl_init_toc_value
dbl_q_tank_volt_limit_warning
dbl_q1_set_limit_min
dbl_q2_set_limit_min
dbl_q3_set_limit_min
dbl_sec_back_to_main
dbl_sec_call_memory
dbl_sec_check_op_hours
dbl_sec_check_p1_counter
dbl_sec_count_op_hours
dbl_sec_data_high_density
dbl_sec_data_low_density
dbl_sec_data_mid_density
dbl_sec_flush_change_cm
dbl_sec_flush_change_cm_edi
dbl_sec_flush_change_pp
dbl_sec_flush_cm_go_idle
dbl_sec_flush_edi_go_idle
dbl_sec_flush_ro_go_idle
dbl_sec_flush_upw_uf
dbl_sec_flush_upw_uf_sv6_close
dbl_sec_flush_upw_uf_sv6_open
dbl_sec_init_flush_upw_no_uf
dbl_sec_init_flush_upw_uf
dbl_sec_init_uf_sv6_close
dbl_sec_init_uf_sv6_open
dbl_sec_io_command_hs
dbl_sec_message_switcher
dbl_sec_no_comm_warning
dbl_sec_no_ps1_over
dbl_sec_no_ps1_sv1_close
dbl_sec_no_ps1_sv1_open
dbl_sec_p2_12_active
dbl_sec_p2_12_sleep
dbl_sec_ps1_error
dbl_sec_ps1_warning
Default value
F
F
F
F
F
F
F
50
5
0,055
0,055
25
3
0,01
1
0,035
0,035
1200
60
600
1800
60
3600
86400
43200
3600
3600
300
0
300
60
300
15
60
300
2700
10
10
30
2
120
10
5
2
300
10800
1800
15
Help
TRUE -> Change all Packs is active
TRUE -> Water production is running
TRUE -> LF1 is in working range
TRUE -> LF2 is in working range
Not used anymore
TRUE -> Tank is full
TRUE -> T3 is over limit
Init value of Q1 Sensor
Init value of Q2 Sensor
Init value of Q3 Sensor
Init value of Q4 Sensor
Init value of Tank Sensor
Init value of TOC (abs (Q3-Q4))
(0,01V) on all sensors -> COM failure!
Q1 < 1µS -> Error
Q2 < 0.035µS -> Error
Q3 < 0.035µS -> Error
After 2 mins back to main
Refresh memory display (1 min)
Update operating hours (10 mins)
Check P1 counter (30 min)
Counter for operating hours (1 min)
Acquire data every 1 hour
Acquire data every 24 hours
Acquire data every 12 hours
Init flushing Con.M. (1 hr)(AQ²E only)
Flushing Con.M. after change (1 hr)
Init flushing PropPak (5 min)
Flush Con.M. when tank full (0 sec)
5 mins flush EDI when Tank full
Flush RO when tank full (1 min)
5 mins flush UF after change
Flushing PW -> SV6 close (15 sec)
Flushing PW -> SV6 open (15 sec)
5 mins first flush UPW-Pack
45 mins first flush UF module
Init flushing PW -> SV6 close (10 sec)
Init flushing PW -> SV6 open (10 sec)
Handshake with control PCB (30 sec)
Switch Messages every (2 sec)
Time to show error NO COMM (2 mins)
PS1 problem is over, 10s wait, back
PS1 problem, close SV1 for 5s
PS1 problem, open SV1 for 2s
Recirculating time (5 min) (when full)
System sleep time (3 hr) (when full)
30 mins Pressure problem -> STOP
15 secs Pressure problem -> Warning
Page 61