Installation and start up
4.3 Hydrogen warning (for Aquinity² E)
During the electro deionization process, small amounts of hydrogen and oxygen gas will be
produced inside the tank. These gases will be drained out of the cell through the product
water stream. During the storage of the product water, the gas mixture may accumulate in
the tank and housing unit, if used, in the case of insufficient air ventilation. This gas mixture
may be flammable if not properly ventilated and if exposed to an ignition source.
The product warning label on the EDI tank opening must remain affixed thereto.
If the water treatment system houses an EDI device, the enclosed product
warning label must be affixed to the outside of the housing container and such
housing container must be properly ventilated. No fire or fire causing devices
should be used near the EDI or its housing container.
Fire and explosion product warning labels are attached to this product and enclosed with
the documentation package for use with external housing containers as described above.

4.4 Setting up the system

The system is intended for dry indoor use only. Please set the system upon an even,
horizontal surface. The surface must be resistant to applied chemicals like disinfectant,
acid, dye, solvents etc. The tank must be located at least at the ground level of the system.

4.5 Sealing plug

In- and outlets of the device are sealed with plugs to prevent leaks during delivery. Before
connecting the system to the water supply and waste water drain, remove the plugs from
the tube connecters on rear side of the device. Lightly depress the fastening ring and gently
remove plug.
membraPure GmbH
Wolfgang-Küntscher-Str.14
16761 Hennigsdorf
Germany
Warning
Potential flammable gas mixture resulting from
-
production of oxygen and hydrogen gases!
Avoid exposure to an ignition source (e.g. open
-
flames, electrical sparks, etc.) when opening the
tank.
Prior to inspection or maintenance work, ensure
-
that the enclosed air volume of the product
storage tank and space in which the EDI is
housed is exchanged using a large excess of oil
free ambient air.
Figure 10: Corresponding content of the warning label
Tel.: +49 (0) 3302/201-20 0
Fax.: +49 (0) 03302/201-20 21
E-Mail: info@membrapure.de
Internet: www.membrapure.de