Service setup
9.1.10 Reset configuration
This tool is used to restore the initial configuration settings of the device.
Requirements: USB-stick with the config.mcf is plugged in the USB-port.
9.1.11 Reset calibration
This tool is used to restore the initial calibration settings of the device.
Requirements: USB-stick with the calib.mcf is plugged in the USB-port
9.1.12 Flushing, disinfection timer
This tool is used to change the time of flushing and disinfection.
Press the button „Reset Config."
-
* Please make sure that there is a USB-stick
(max. 4 GB, in FAT-format) connected to the USB-port!
-> copy files from \Hard Disk\... into \Nandflash\...
This is a tool to load configuration settings of the
-
device (all configuration-parameters are not
adjustable directly in the software)
See chapter 5.2.5 about exporting data
Press the button „Reset Calib."
-
* Please make sure that there is a USB-stick
(max. 4 GB, in FAT-format) connected to the USB-port!
-> copy files from \Hard Disk\... into \Nandflash\...
This is a tool to load calibration settings of the
-
device (most calibration-parameters are adjustable
in the software)
See chapter 5.2.5 about exporting data
Press the button „Flush/Dis. Timer"
-
Page 54