System Control Panel
(13) No Pressure,
check pressure sensor!
(15) Warning: T3 is
over 40°C!
(16) Tanksensor
overflow! Check
tanklevel sensor!
(17) Warning: Q3 is
over limit!
(18) OPERATING
HOUR LIMIT ProPak
R10, Exchange
necessary
(32) Please Disinfect
System!
membraPure GmbH
Wolfgang-Küntscher-Str.14
16761 Hennigsdorf
Germany
No feed water pressure.
System temperature caused
by UV lamp or recirculation
pump too high.
Product water tank
overflowing.
Q3 exceeds the limit.
Pretreatment module
ProPak-R10 exceeded the
maximum operating hours.
Maximum time until the next
required disinfection has
expired.
Table 5: Display warning messages
Tel.: +49 (0) 3302/201-20 0
Fax.: +49 (0) 03302/201-20 21
E-Mail: info@membrapure.de
Internet: www.membrapure.de
system for leaks. Check the
function of the RO process.
Check the main water supply.
Check the pressure sensor.
Inform the customer service.
Check Interval-Mode. Check the
temperature of the environment
and the feed water.
Check the Tank-settings. Check
the tanklevel sensor.
Replace the polisher module
UPW-Pak.
Reset the expired operating
hours.
Replace the pretreatment module
ProPak-R10. Reset the expired
operating hours.
Disinfect the system.