Replacement Of Ultrafiltration Module; Disinfection (Uf-Systems Only) - MP Aquinity 2 E Service Manual

Table of Contents

Advertisement

Care and maintenance
The system is operational.
-
Check the UV-submersible-lamp for leakages.
-
Turn on the submersible lamp by plugging the power adapter into an electrical
-
outlet.

6.6 Replacement of ultrafiltration module

For exchanging of ultrafiltration module follow the steps listed in chapter 5.2.1
-
Caution by removing tubes from the ultrafiltration module (water leakage).
-

6.7 Disinfection (UF-systems only)

To perform a disinfection please use the disinfection-tool. To do that please follow the
steps in chapter 5.2.7. If the disinfection-tool is started, it must be finished. It is impossible
to abort the tool. Instructions for installing the disinfection-cell are listed below.
As disinfectant ClO
installing / removing the disinfection cell.
Unplug the tubes out of the polisher module UPW-Pak and put them into the
-
connections of the disinfection cell
membraPure GmbH
Wolfgang-Küntscher-Str.14
16761 Hennigsdorf
Germany
is used. Basically wear googles and latex gloves when
2
Figure 19: Mounted disinfection cell
Tel.: +49 (0) 3302/201-20 0
Fax.: +49 (0) 03302/201-20 21
E-Mail: info@membrapure.de
Internet: www.membrapure.de
page 41