Service setup
The scale value is set in factory and should not be changed. The adjustment is
done by using the offset normally. For more detailed calibration instructions
please see chapter 9.2.1.

9.1.2 Limit setting of conductivity sensors

9.1.3 Tank setup

Press the button "Q-" or "Q+" to change the scale-
-
value of the conductivity measurement (upper value)
or the offset (lower value)
Press the button "T+" or "T-" to change the scale-
-
value of the temperature measurement (upper value)
or the offset (lower value)
Entries change is saved automatically
-
Press the button „Q-Limits"
-
Press the button "-" or "+" to shift the limit of
-
conductivity sensors in µS/cm
Note: The unit of these conductivity limits is always
shown in µS/cm
Press the button „Tank Setup"
-
Press the button "V+" or "V-" to change the maximum
-
voltage range of the tank level sensor
Press the button "VOL+" or "VOL-" to change the
-
maximum scale-range of the tank level indicator.
Entries change is saved automatically
-
Page 50