Certificate of Analysis (CoA) are matching and correct. 1. Introduction Thank you for purchasing the Shodex product. Shodex CLNpak EV series is a GPC clean-up column packed with a styrene divinylbenzene copolymer. It is suitable for clean-up of residual pesticide components in foods.
Page 2
Thus, dilution of such samples with eluent is recommended. The approximate maximum lipid load per injection is less than 180 mg. Note · Use of guard column is recommended to protect the analytical column. SHOWA DENKO K.K. (https://www.shodex.com/)
Page 3
(2) Check miscibility/solubility of the desired new solvent and the solvent currently filled in the column. (3) When replacing the current solvent with a miscible solvent, first introduce 1:1 mixture of the current solvent and the new solvent, and then replace it with 100 % of the new solvent. SHOWA DENKO K.K. (https://www.shodex.com/)
Page 4
9. Column Inspection Inspection method is described in the Certificate of Analysis (CoA). Theoretical Plate Number (N) and Asymmetry Factor (Fas) were calculated using the below equations. (1) Theoretical Plate Number (N) (2) Asymmetry Factor (Fas) SHOWA DENKO K.K. (https://www.shodex.com/)
Page 5
(2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex...
Page 6
11. Warranty (Ver. 4COLUMN) (1) Showa Denko K. K. warrants that the Shodex product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.