Step Description
10
•
Insert the NG11 filter into the cell holder. Make
sure to install the filter in the correct orientation.
•
Install the instrument cup.
Note: The filter is light-tight. Therefore, it is not
necessary to install the instrument cap.
•
Push
to measure the absorbance.
•
Make a note of the measured value of the
NG11 filter and enter the actual value in the
Certificate of Service Inspection.
11
•
Insert the NG5 filter into the cell holder. Make
sure to install the filter in the correct orientation.
•
Install the instrument cup.
Note: The filter is light-tight. Therefore, it is not
necessary to install the instrument cap.
•
Push
to measure the absorbance.
•
Make a note of the measured value of the
NG5 filter and enter the actual value in the
Certificate of Service Inspection.
Picture
Inspection
31