  • Page 1 User manual 2015-11-23 ® CuroCell AREA Zone C5 Self-adjusting pressure care mattress User manual item number: 95-001308...
  • Page 2: Table Of Contents

    Contents Important safety precautions Introduction Product description Instruction for use Cleaning and reconditioning Storage Troubleshooting Technical description Spare parts Accessories Warranty Other information Own notes...
  • Page 3: Important Safety Precautions

    The product may not be combined, assembled or repaired with other parts, accessories or spare parts other than those described in this manual or other documentation from Care of Sweden. The mattress are protected with a hygiene cover; Please try to avoid the simultaneous use of multiple hygiene cover with this mattress.
  • Page 4: Introduction

    We would be very grateful for all information involving accidents or incidents relating to our products, by reporting them immediately to us, here at Care of Sweden. The system has been tested and approved according to the following standards:...
  • Page 5 Mattress guide This practical guide gives you as a care giver, help in choosing the correct mattress from the Care of Swedens range. CuroCell® AREA C5 is a mattress in function group B. This information can be found on our website www.careofsweden.com or you can order a copy from our Customer services by sending your details to info@careofsweden.se.
  • Page 6 Down the stairs • Walk backwards until the main part of the mattress has slide pass the top stair. • Turn the back to the patient and look in the moving direction. • Hold one hand on the stair handle and pull with the other hand on the middle handle of the mattress.
  • Page 7: Product Description

    Available sizes are (width, length and height) 80/85/90/105/120 x 200/210 x 18 cm. For request of other sizes please contact Care of Sweden or your local dealer. 2.1 Description cover CuroCell® AREA Zone C5 is delivered with a hygiene cover. It is easy to handle and maintain and manage high standards of cleaning and hygiene.
  • Page 8: Instruction For Use

    3. Instruction for use 3.1 Unpacking the mattress The mattress is packed in a box. 1. Start with making sure that the package is not damaged. When opening the box and any other protective material, avoid using a knife since this could damage the mattress and cover.
  • Page 9: Cleaning And Reconditioning

    4. Cleaning and reconditioning It is important to follow the instructions below before using the product for a different user. Consult the hygiene responsible or the supplier for help and instructions when unsure. NOTE! Check the hygiene cover each time the product is cleaned. If damaged, it must be replaced or repaired.
  • Page 10: Storage

    The patient is bottoming out Open covers and check cells and tubing. Check again through Hand check (see chapter 3.2). If your questions are not answered by information above, please contact Care of Sweden or your local dealer. 7. Technical description 7.1 Technical specification...
  • Page 11 7.3 Symbol Definitions XXXX-XX-XX Article No Manufacturer Year-Month-Day Foot placement Read the user manual Group of function CE marked acc. to Recycling MDD 93/42/EEC Recommended user weight Directly on bed base Not in domestic refuse, Patient information follow recycling instructions –...
  • Page 12: Spare Parts

    120x220 cm Art. no. Description Size CC-900050 Air cell AREA, 1 pcs 63x15x12 cm If other spare parts are needed, please contact Care of Sweden or your local dealer for further assistance. 9. Accessories 9.1 Accessories Inner Cover AREA Art. No.
  • Page 13: Warranty

    105-120 cm 9.2 Other accessories To ensure the best level of comfort, we recommend use of accessories from Care of Sweden since our products are designed to work well together. Please contact Care of Sweden or your local dealer for accessories or more information on our products.
  • Page 14: Own Notes

    12. Own notes...
  • Page 16 SE-514 23 Tranemo SWEDEN Visit address: Fabriksgatan 5A SE-514 33 Tranemo SWEDEN Cargo address: Byns väg 4A SE-514 33 Tranemo SWEDEN Supplied by Care of Sweden and CuroCell are trademarks of Care of Sweden AB © Care of Sweden AB, 2015...