5. Commissioning, operation
5.1 Mechanical mounting
The sealing faces at the instrument and at the
measuring location have to be undamaged and
EN
clean.
When screwing in the instrument, the force
required to do this must not be applied through
the case, but only through the spanner flats
provided for this purpose and using a suitable
tool.
The correct torque depends on the dimensions of the process connection and the sealing used
(form/material).
Do not overwind the process connection, do not exceed the permissible torques.
When screwing in, do not cross the threads.
For information on tapped holes and welding sockets, see Technical information IN 00.14 at
www.wika.com.
12
WIKA operating instructions digital pressure gauge, model DG-10
Spanner flats