3.COMP
The acoustic warning signal and the error message
(Error 58 TILT) can be suppressed if the total station is
not required to be properly levelled up. (eg, on swaying
platforms or ships).
For normal use set COMP = ON.
When switching the total station on the function is
automatically set to COMP = ON.
Transverse
Longitudinal
vertical line
vertical line
standing axis
standing axis
When the compensator is switched off (COMP = OFF)
the vertical angle refers to the standing axis.
When the compensator is switched on (COMP = ON)
the vertical angle refers to the vertical line (plumb line).
4.Hzcomp (dual axis compensation)
Function HZCOMP = ON corrects the Hz-angles for
the tilt of the standing axis.
Switching the compensator (COMP) off also sets the
function HZCOMP to OFF.
When switching the total station on the function is
automatically set to HZCOMP = ON.