Set The Calibrator Ip Address, Port And Calibrator Number (Ipcal); Set The Calibration And Validation Average (Calavg); Set Enable Calz Before Validation (Valzo) - Scanivalve MPS4232 Hardware And Software Manual

Miniature pressure scanner
Table of Contents

Advertisement

Section 5: Software
RETURNS
EXAMPLE
DEFAULT
RANGE

Set the Calibrator IP Address, Port and Calibrator Number (IPCAL)

COMMAND
TYPE
DESCRIPTION
SYNTAX
RETURNS
EXAMPLE
DEFAULT
RANGE
NOTES

Set the Calibration and Validation Average (CALAVG)

COMMAND
TYPE
DESCRIPTION
SYNTAX
RETURNS
EXAMPLE
DEFAULT
RANGE

Set Enable CALZ Before validation (VALZO)

COMMAND
TYPE
DESCRIPTION
SYNTAX
RETURNS
Prompt when ready to accept a new command
SET DELAY 120 45 0 200
>
120 0 0 300
These values are not range checked.
IPCAL
Set/Get
Sets the IP address and port for the calibrator. This is only applicable when using a Scanivalve
SPC4000 or SPC4050 calibrator.
SET IPCAL <ipadd> <port> <calibrator number>
Where:
<ipadd> = Calibrator IP address.
<port> = Calibrator port.
<calibrator number> = Calibrator number.
Prompt when ready to accept a new command
SET IPCAL 10.0.0.122 23 1
>
10.0.0.61 23 1
These values are not range checked.
When 0.0.0.0 0 0 is used, a simulated calibrator for debug is used.
CALAVG
Set/Get
Sets the number of samples taken from the MPS for a calibration and validation.
SET CALAVG <calrate> <calavg>
Where:
<calrate> = Is the scan rate at which the samples are taken for the calibration/validation.
<calavg> = Is the number of sample taken for the calibration/validation. Valid values range
between 1 and 32000.
Prompt when ready to accept a new command
SET CALAVG 12 128
>
0.9 16
<0 to 850> <1 to 32000>
VALZO
Set/Get
When set (1) a CALZ is performed at the start of each temperature during a validation.
SET VALZO <0 or 1>
Prompt when ready to accept a new command
64
MPS4232