Initial Use and Operation

When started, the aspirator displays the following:
The startup configuration while performing it's self-check procedure.
The LED alternately blinks yellow and green. The alternating yellow/green pattern indicates
that the device is ready to operate and that user input is required to begin operation.
When the self-check procedure completes, the aspirator the aspirator is operational.
If you want to operate the aspirator in Continuous mode with the initial Adult settings and
Smart Flow enabled, follow these steps.
1. Press the parameter button that corresponds to the procedure you wish to perform.
2.
Adjust the vacuum level for your use, use the Dial to adjust it, then press the Confirm button.
Test the aspirator to verify the vacuum level.
3.
If needed, calibrate Smart Flow. See, "Smart Flow Calibration", in this chapter.
4.
Note:
Refer to, "Continuous Mode Settings", in this chapter for descriptions of these procedures.
Note:
Note:
Refer to section, "Suction Settings" in this chapter for additional information and instructions
to change modes, choose procedures, and set vacuum levels.

Suction Settings

Aspirator settings include:
Suction Mode
Continuous Mode parameters include:
Intermittent Mode parameters include:
The sections that follow describe these settings and provide instructions for choosing a value.
906-0330-01-01 Rev. D
Surgical
Pharyngeal
Tracheal
CASS
When Smart Flow is ON, the pump is operating at a low speed. Occlude the suction
tubing to ensure that the collection system is intact and that the set vacuum level is
reached.
Failure to adjust a highlighted parameter or press the Confirm button within 5 seconds
automatically cancels the action. This system ensures inadvertent button presses do not
result in an unintended change
Upon initial start, Smart Flow is ON (enabled). For information about the
Smart Flow feature, see, "About Smart Flow", later in this chapter.
Continuous
Intermittent
Suction Procedure (Surgical, Pharyngeal, Tracheal, CASS)
Vacuum level
Smart Flow
Smart Flow calibration
Vacuum level
ON Time
OFF Time
ZOLL 330 Multifunction Aspirator Operator's Guide
Using the Aspirator User Interface
4-7