Standby
To conserve power, reduce noise and allow for immediate us of the aspirator, the aspirator has a
"Standby" feature.
To start and stop Standby, press the Standby button. When Standby is active:
Standby displays in the Alarm Message Center (AMC)
•
The vacuum pump is de-energized, not flowing (zero RPM)
•
The LCD is dimmed
•
The LED array alternates Green/Yellow to indicate the device status
•
When standby is stopped, by pressing the Standby button, the aspirator immediately begins
operation with the last settings.
Pause
To minimize risk to the patient, the device enters a Pause state where the device waits for input
form the user.
Pause displays in the Alarm Message Center (AMC)
•
The vacuum pump is de-energized, not flowing (zero RPM)
•
The bleed valve relieves pressure
•
The LED array alternates Green/Yellow to indicate the device status
•
Pause occurs whenever the device switches between modes (Continuous and Intermittent) to
prevent patient exposure to an unintended vacuum level. The Pause state ends as soon as the
user enters a vacuum level.
906-0330-01-01 Rev. D
ZOLL 330 Multifunction Aspirator Operator's Guide
Product Overview
2-11