4

Measure

WARNING
If the Vivatmo pro device has been stored in warm and humid environment or not been in use for a longer
time, reconditioning may be required and starts automatically during first measurement. Therefore per-
form a test measurement every day before using the device with the patient, see "7.3 Reconditioning".
Following conditions can influence correct measurement results and shall be avoided:
• Smoking or tobacco consumption for at least 1 hour before the measurement.
• Eating or drinking at least 1 hour before the measurement, especially nitrate rich food
(e.g. spinach).
• Strenuous exercise.
• Rooms with high air pollution, high ambient NO or open flames, for example gas range, smoke,
or from tobacco consumption.
Disposable mouthpiece shall be handled with care for correct measurement results:
• Use only disposable mouthpieces approved for this device and from intact and unopened
pouches that are not expired. See the expiration date on the pouch. Store in outer package.
• Do not clean the disposable mouthpiece.
• Take care to use regular disposable mouthpieces with transparent cover. The use of Level 0
disposable mouthpieces with white cover and a "0" on the disposable pouch leads to 0 ppb
measurements.
Always check your Vivatmo pro basestation, handheld, and the disposable mouthpiece for damages
before using to avoid injury or malfunction.
Leakage at the disposable fitting may lead to a lower measurement result.
A mouthpiece can be used for one patient measurement only. In case of an unsuccessful
measurement note that at most 5 attempts possible with 1 mouthpiece. Use the mouthpiece within
15 minutes after opening its pouch.