6.5.5 Release the handheld

In case another handheld was installed before, it is disconnected by tapping the Release Handheld button. If you want
dispose of the old handheld, clear the memory and take out the battery.
54
© Bosch Healthcare Solutions GmbH
1. Keep the ENTER button and the Arrow right button pressed for 3 seconds.
The Clear memory icon is selected.
2. Use the Arrow buttons to select the check mark.
3. Press the ENTER button to confirm the clear memory.
An animation is shown while the Clear memory icon flashes.
4. Press the ENTER button to return to the main screen.
5. Switch off the handheld.