Test Procedure
Test
2.1
2.2
9650-0308-01 Rev. A
Action
Turn on the AED Plus unit.
Press the Treatment button
(first shock).
ZOLL AED Plus Service Manual
Preventive Maintenance Test
Expected Result(s)
!
UNIT OK
!
ADULT PADS
!
DON'T TOUCH PATIENT,
ANALYZING
!
TREATMENT ADVISED
(or SHOCK ADVISED)
!
DON'T TOUCH PATIENT.
!
PRESS TREATMENT BUTTON
(or PRESS SHOCK BUTTON)
• Analyzer indicates a defibrillator
discharge at the configured
first-shock energy level (120, 150
or 200 joules ±15%).
!
TREATMENT DELIVERED
(or SHOCK DELIVERED)
• An updated shock count appears
on the display (SHOCKS: 1).
!
DON'T TOUCH PATIENT,
ANALYZING
!
TREATMENT ADVISED
(or SHOCK ADVISED)
!
DON'T TOUCH PATIENT.
!
PRESS TREATMENT BUTTON
(or PRESS SHOCK BUTTON)
Pass
Fail
!
!
!
!
1–5