5. Troubleshooting
5. Troubleshooting
5. Troubleshooting
5. Troubleshooting
Error in Motor Rotation
No.
Check error in motor.
Measure the resistance of each
case in probe Connector
1) Pin number 5 and 8
1
2) Pin number 10 and 11
3) Pin number 13 and 15
4) Pin number 16 and 17
2
In case of normal resistance.
www.mcubetech.co.kr
Check point
Description
Normal resistance value
15 +/- 1 (Ohms)
Retry after replacing analog
board.
-50-