5. Troubleshooting
5. Troubleshooting
5. Troubleshooting
5. Troubleshooting
System is not turned on
No.
Is POWER key pressed for more than 1
1
second?
2
Voltage of battery module.
3
Is battery module connected?
Connector between the Key Board and
4
the Control board ?
Is connected from the Battery Module to
5
the Control Board in the right direction?
6
Not solved yet?
www.mcubetech.co.kr
Check point
Description
The system operates when pressing the
POWER key for more than 1 second.
Above +7V
If not, connect the module.
Connect the connector between two
boards.
Control Board – Analog Board
Analog Board – Charge Board
Charge Board – Battery Board
Exchange control boards.
-40-