Mcube CubeScan BioCon-500 Service Training page 18

Hide thumbs Also See for CubeScan BioCon-500:

Advertisement

3. System Setup
3. System Setup
3. System Setup
3. System Setup
Calibration
1. Check the error in the ultrasound transducer.
2. Verify the start position of the plane motor.
3. Verify the direction of rotation of the plane motor.
4. Check the plane motor rotation.
5. Set the angle motor offset.
6. Check the angle motor rotation.
7. Set the reference threshold.
www.mcubetech.co.kr
-18-