5. Troubleshooting
5. Troubleshooting
5. Troubleshooting
5. Troubleshooting
Initial screen is not display after turn-on
No.
1
Is backlight on in LCD?
Is the FPC connector to the LCD module
2
well connected?
Check the voltage between the Control
3
board ground and 12
Check the voltage between the Control
4
board ground and 13
5
Is the system turned off?
6
The system is not turned off.
www.mcubetech.co.kr
Check point
th
pin of JP5.
th
pin of JP5.
Description
Check if the connector for backlight
of LCD is connected.
Check if the connector is connected
properly.
Check if -23V (+/-1V).
Check if -18V (+/-1V).
Retry after replacing the LCD module.
Test again after replacing the Control
board.