Pyxis Lab, Inc. Standard Limited Warranty Pyxis Lab warrants its products for defects in materials and workmanship. Pyxis Lab will, at its option, repair or replace instrument components that prove to be defective with new or remanufactured components (i.e., equivalent to new).
1 Introduction The Pyxis SP-200 Handheld Colorimetric Analyzer is a single- and multi-parameter colorimetric testing plat- form. With the use of auxiliary colorimetric reagent kits, the SP-200 can quickly test a variety of water quality parameters on the spot with high sensitivity.
• Bluetooth/USB Adapter for Desktop P/N: MA-NEB • User Manual available online at https://pyxis-lab.com/support/ 3.2 Optional Accessories The following optional accessories can be ordered from Pyxis Customer Service (order@pyxis-lab.com) or Pyxis E-Store at https://pyxis-lab.com/shop/. Table 2. Optional Accessories Accessory Name Part Number (P/N)
4. Replace the battery compartment cover and ensure that the sealing O-ring is lying flat on the battery holder. 5. Fasten the two screws. NOTE Failure to properly seat the O-ring may result in water damage to the SP-200. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
The SP-200 turns itself off after 60 seconds without user interaction detected. This is done to conserve battery life. NOTE This auto-time off setting may be customized by the user as desired through the uPyxis® Mobile or Desktop App. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
6.1 Colorimetric Methods A wide range of colorimetric methods is supported by the SP-200 analyzer and the number of them keeps increasing with continuous development of Pyxis. The following table is list of Supported Colorimetric Meth- ods: Table 3. Supported Colorimetric Methods...
The rate of the reaction is often faster than the standard preset time. If the concentration remains stable on the concentration-time plot, press Stop ( ) to terminate the timing step early. The last concentration measurement will be displayed. Figure 5. Figure 6. Figure 7. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
) to launch the Choose Unit screen (Figure 8) or the Choose Form screen (Figure 9). 4. Press » ( ) to navigate to the desired parameter option. 5. Press OK ( ) to select the desired parameter option. Figure 8. Figure 9. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
7. Press OK ( ) to start the low calibration. 8. The message “Low point calibration succeeded” will appear on the display (Figure 12). 9. Press to exit. Figure 10. Figure 11. Figure 12. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
7. Press OK ( ) to start the slope calibration. 8. The message “High point calibration succeeded” will appear on the display (Figure 15). 9. Press to exit. Figure 13. Figure 14. Figure 15. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
). The display updates as shown in Figure 16. 4. Press OK ( ) to start the factory default of the desired method. 5. The message “Restore data succeeded” will appear on the display (Figure 17). Figure 16. Figure 17. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
) to launch the Diagnosis screen where raw measurement data are displayed (Figure 19). The information has no use for normal operation, but instead is used for device troubleshooting. Provide an image of both the Information screen and the Diagnosis screen when you contact Pyxis (service@pyxis- lab.com) for troubleshooting your device or call +1 (866) 203-8397.
). The display updates as shown in Figure 20. 6. Press OK ( ) to start the factory reset. 7. The message “Data recovery is successful!” will appear on the display (Figure 21). Figure 20. Figure 21. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
9 Use with uPyxis® Mobile App 9.1 Download uPyxis® Mobile App Download uPyxis® Mobile App from Apple App Store Google Play. Figure 22. uPyxis® Mobile App installation SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
1. Follow the steps in the Bluetooth Connection to Devices section to make the SP-200 discoverable. 2. Open uPyxis® Mobile App. 3. On uPyxis® Mobile App, pull down to refresh the list of available Pyxis devices. 4. If the connection is successful, the SP-200 and its Serial Number (SN) will be displayed (Figure 26).
9.3 Setting Screen From the Setting screen, the user can set the Power off time and Screen off time in seconds. Figure 24. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
(approximately 10 ft without obstructions) for the entire duration of the up- date. Once the update is complete, the SP-200 will reboot which will disconnect the SP-200 from the uPyxis® Mobile App. Figure 25. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
10 Use with uPyxis® Desktop App 10.1 Install uPyxis® Desktop App Download the latest version of uPyxis® Desktop software package from: https://pyxis-lab.com/upyxis/ this setup package will download and install the Microsoft.Net Framework 4.5 (if not previously installed on the PC), the USB driver for the USB-Bluetooth adapter (MA-NEB), the USB-RS485 adapter (MA-485), and the main uPyxis®...
4. On uPyxis® Desktop App, click Device Connect via USB-Bluetooth (Figure 30). 5. If the connection is successful, the SP-200 and its Serial Number (SN) will be displayed in the left pane of the uPyxis® window. Figure 28. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
Once connected to the device, a picture of the device will appear on the top-left corner of the window and the uPyxis® Desktop App will default to the System screen. From the System screen, users can upgrade the firmware by selecting an appropriate firmware file (contact service@pyxis-lab.com for these firmware Upgrade Firmware.
Read Datalog, deleted by clicking Delete, or exported by clicking Export as .CSV File. Figure 31. SP-200 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
Page 26
150 °F in summer and -20 °F in winter. Remove batteries from the SP-200 battery compartment when in storage for a long period time. Do not leave a sample vial in the sample vial compartment. 12 Contact Us Pyxis Lab, Inc 1729 Majestic Dr. Suite 5 Lafayette, CO 80026 USA www.pyxis-lab.com...