Pyxis SP-400 User Manual

Pyxis SP-400 User Manual

Handheld fluorometer
Hide thumbs Also See for SP-400:

Advertisement

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SP-400 and is the answer not in the manual?

Questions and answers

Summary of Contents for Pyxis SP-400

  • Page 2 SP-400 Handheld Fluorometer User Manual August 10, 2020 Rev. 1.31 Pyxis Lab, Inc. 1729 Majestic Dr. Suite 5 Lafayette, CO 80026 USA www.pyxis-lab.com © 2017 Pyxis Lab, Inc. Pyxis Lab Proprietary and Confidential...
  • Page 3: Table Of Contents

    Table of Contents 1 Introduction 1.1 Main Features ......... 2 Specifications 3 Unpacking Instrument 3.1 Standard Accessories .
  • Page 4 11.1 Maintenance Best Practices and Quick Tips ......11.2 Methods to Cleaning the SP-400 .......
  • Page 5 Pyxis Lab, Inc. Standard Limited Warranty Pyxis Lab warrants its products for defects in materials and workmanship. Pyxis Lab will, at its option, repair or replace instrument components that prove to be defective with new or remanufactured components (i.e., equivalent to new).
  • Page 6: Introduction

    1 Introduction The Pyxis SP-400 is a uniquely designed handheld multimeter that measures both PTSA and conductivity, simultaneously. It is a cuvette-less device. Less than 5 mL water sample is needed to fill the sample cell for proper measurement. 1.1 Main Features The SP-400 includes the following features: •...
  • Page 7 Calibration Solutions 500 µS/cm, 1000 µS/cm, 2500 µS/cm, 5000 µS/cm, combined 1000 µS/cm and 100 ppb PTSA, and user-defined (500–5000 µS/cm) With Pyxis’s continuous improvement policy, these specifications are subject to change without notice. Table 3. Temperature Specifications Item Specification Range 42–112 °F (5–45 °C)
  • Page 8: Unpacking Instrument

    The SP-400 is powered by four alkaline batteries. Typical battery life lasts for 10,000 measurements or 10 months. When the battery capacity is critically low, the SP-400 will display a ”LOW BATTERY” warning for 5 seconds and then automatically turn off.
  • Page 9: Instrument Overview

    5. Fasten the two screws. NOTE Failure to properly seat the O-ring may result in water damage to the SP-400. 5 Instrument Overview 5.1 Control Keys The SP-400 has three control keys, as shown in Figure 2. The left (...
  • Page 10: Main Module On/Off

    To turn off the SP-400: Press and hold for about three seconds. Release when the LCD display turns off. The SP-400 turns itself off after 60 seconds without user interaction detected. This is done to conserve battery life. NOTE This auto-time off setting may be customized by the user as desired through the uPyxis®...
  • Page 11: Conductivity Temperature Compensation

    6.3 High Color and Turbidity Warning The SP-400 has extra channels to measure sample turbidity and color to automatically compensate sam- ple color and turbidity interference. If sample turbidity and color values determined are too high, a PTSA measurement warning will be displayed. In such a case, the user should filter the sample for PTSA mea- surement.
  • Page 12: Calibration

    1412 (or 1413) µS/cm standard, to calibrate the SP-400. For convenience, the SP-400 can be calibrated using a combined standard with 100 ppb PTSA + 1000 µS/cm KCl conductivity (PTSA-1010). It is highly recommended to use the PTSA standalone calibration procedure if the user desires to achieve higher accuracy for low range PTSA measurements (<...
  • Page 13 Figure 3. Figure 4. Figure 5. Figure 6. SP-400 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
  • Page 14: Conductivity Calibration (500, 1000, 2500, Or 5000 Μs/Cm)

    ) to return to the CONDUCTIVITY CALIBRATION screen, or (c) Press Exit ( ) to abandon calibration entirely. 8. Once calibration begins, the SP-400 reads the sample and displays the value in the Measured section. A slight variance from the target is acceptable. 9. The message “Calibration...
  • Page 15: User-Defined Conductivity Calibration

    ) to return to the USER DEFINED CALIBRATION screen, or (c) Press Exit ( ) to abandon calibration entirely. 9. Once calibration begins, the SP-400 reads the sample and displays the value in the Measured section. A slight variance from the target is acceptable. 10. The message “Calibration...
  • Page 16 Figure 11. Figure 12. Figure 13. Figure 14. Figure 15. SP-400 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
  • Page 17: Combined Ptsa And Conductivity Calibration

    7.4 Combined PTSA and Conductivity Calibration The following steps will calibrate both conductivity at 1000 µS/cm and PTSA at 100 ppb using Pyxis 100 ppb PTSA + 1000 µS/cm (KCl) Combined Standard (P/N: 21004): 1. Rinse the sample cell three times with the Combined Standard. Fill the sample cell with the Combined Standard.
  • Page 18: Device Information And Diagnosis

    (Figure 21). The information has no use for normal operation, but instead is used for device troubleshooting. Provide an image of both the DEVICE INFORMATION screen and the SYSTEM DIAGNOSIS screen when you contact Pyxis (service@pyxis-lab.com) for troubleshooting your device or call +1 (866) 203-8397. Figure 20.
  • Page 19: Bluetooth Connection To Devices

    8.1.1 Calibrate an ST-500 Series Probe with the SP-400 via Bluetooth The SP-400 can be used to verify the result of an inline Pyxis ST-500 Series probe by measuring the sample water taken from the inline probe sample line. The SP-400 can then be used to calibrate the inline probes over the Bluetooth connection.
  • Page 20 13. The SP-400 will take the probe PTSA measurement three times to verify the calibration (Figure 26). NOTE It takes about one minute for the probe to approach the calibrated reading and the three verifying readings may not be exactly the same as the value measured by the SP-400Ṗ...
  • Page 21: Factory Reset

    8.2 Factory Reset Use the following steps to restore all device parameters to factory default: 1. Power on the SP-400 by pressing 2. Press System ( ) to launch the DEVICE INFORMATION screen. 3. Press Diagnosis ( ) to launch the SYSTEM DIAGNOSIS screen.
  • Page 22: Use With Upyxis® Mobile App

    Mobile App. Ensure that the SP-400 is discoverable by following the steps in the Bluetooth Connection to Devices section. Once the app is open, the app will start to search for the SP-400. When the uPyxis® Mobile App connects to the SP-400, then press on the SP-400.
  • Page 23: Setting Screen

    When connected, the uPyxis® Mobile App will default to the Setting screen. From the Setting screen, the user can set the Power off time and Screen off time in seconds. Figure 32. SP-400 User Manual service@pyxis-lab.com | +1 (866) 203-8397...
  • Page 24: System Screen

    The firmware update process takes some time and will require the SP-400 to stay within range (approximately 10 ft without obstructions) for the entire duration of the up- date. Once the update is complete, the SP-400 will reboot which will disconnect the SP-400 from the uPyxis® Mobile App. Figure 33.
  • Page 25: Use With Upyxis® Desktop App

    10.2 Connecting to uPyxis® Desktop App Ensure that the SP-400 is discoverable by following the steps in the Bluetooth Connection to Devices sec- tion. Once the app is open, to find your device, click on Device, then Connect via Bluetooth..
  • Page 26: System Screen

    The firmware update process takes some time and will require the SP-400 to stay within range (approximately 10 ft without obstructions) for the entire duration of the up- date. Once the update is complete, the SP-400 will reboot which will disconnect the SP-400 from the uPyxis® Mobile App. Figure 36.
  • Page 27: Datalog Screen

    10.4 Datalog Screen From the Datalog screen, the user can view, delete, and export the internal log files of the SP-400 by clicking Read Datalog List and selecting the desired datalog (these are separated by month). The SP-400 will then...
  • Page 28 11 Device Maintenance and Precaution 11.1 Maintenance Best Practices and Quick Tips For greatly increased working life and ease of use of the SP-400 follow the list of maintenance best practices and quick tips below: • Rinse the sample cell at least three times with the desired solution before turning on.
  • Page 29 11.3 Storage Do not expose the SP-400 to an extreme high or low temperature condition such as leaving the SP-400 inside an unattended automobile. 12 Contact Us Pyxis Lab, Inc 1729 Majestic Dr. Suite 5 Lafayette, CO 80026 USA www.pyxis-lab.com...

Table of Contents