Light Shield Cover ....................8 Start The SP-910..................10 Battery Installation ....................10 Description of the Navigational Control Pad ............10 Turning on the SP-910 ..................11 Main Page ......................11 Turning Off The SP-910 ..................11 The SP-910 Auto Power Off.................. 12 Auto LCD Power Saving ..................
Page 4
Operation ......................27 Turbidity Calibration ..................... 27 Absorbance Measurement ..................30 Update SP-910 firmware via Bluetooth Interface ..........31 Download and Install the Latest Nebula Software ..........31 Turn on SP-910 ..................... 32 Connect Nebula to the SP-910 ................32 Upgrade the ChemTreat Product List ............
Page 6
Tables Table 1 Feature Groups on Main Menu ............... 11 Table 2 List of Supported Colorimetric Methods ..........18 Table 3 Wavelength of each channel ..............30 Table 4 Contact Information ................38...
Page 7
A Repair Authorization Number (RA) must be obtained from the Technical Support (service@pyxis- lab.com) before any product can be returned to the factory. Pyxis will pay freight charges to ship replacement or repaired products to the customer. The customer shall pay freight charges for returning products to Pyxis.
1. Specifications are subject to change without notice for continuous development. 2. The fluorescein range in earlier versions of the SP-910 may be only up to 20 ppb. To extend the upper limit to 600 ppb, please contact Pyxis customer support at...
The SP-910 is pre-calibrated for measuring fluorescein in the range of 0 to 600 ppb. The SP-910 is pre-calibrated for measuring turbidity in the range of 0 to 200 NTU. Automatically select the primary wavelength according to the method selected and switches to the secondary wavelength to extend the primary measurement range.
The light shield cover should be in the closed position during storage, transportation, and measurements, especially during the turbidity and fluorescence measurements. When turned on, The SP-910 carries out self-diagnosis including checking the performance of a variety of optical devices. The light shield door shall be at the closed position to shield interference from ambient light during self-diagnosis.
Figure 2 Open and Close the Light Shield Cover Warning Magnetic sensitive devices, including but not limited to, credit cards, watches, hard disks, should be keep at a distance of at least 2 inches from the Light Shield Door to avoid possible damage and/or loss of information recorded.
LOW BATTERY warning. Replace all four batteries to resume operation of the SP-910 after the battery warning. The SP-910 battery compartment, shown in Figure 3, is on the back side of the instrument. Insert a small pad underneath the screen area to make the back surface level when the instrument is turned upside down.
After new batteries installation, the SP-910 will not be automatically turned on. To turn on SP-910, press the OK key, and release the OK key when the LCD is lit. You can navigate the main page menu and launch an operation by pressing on an icon.
Turn the SP-910 off by navigating to Power icon and press the OK key. Alternatively, you can turn off the SP-910 by pressing OK key for 5 seconds in any menu. 2.6 The SP-910 Auto Power Off The SP-910 automatically turns itself off with no-key activity for a given period, except for during a measurement.
2. Place the sample vial into the sample vial compartment and slide the light shield cover to the closed position. 3. Press the M-P icon on the main page. The SP-910 will start to measure the PTSA concentration. 4. The SP-910 will display the ppm concentration of the last selected ChemTreat cooling product.
During the fluorescence measurement to determine the PTSA concentration, the SP- 910 checks the sample turbidity. If the sample turbidity value detected is greater than 40 NTU, The SP-910 will display a warning. For best results, the sample should be filtered if turbidity exceeds 40 NTU.
2. Place the sample vial into the sample vial compartment and slide the light shield cover to the closed position. 3. Press the M-F on the main page, then press the OK button, the SP-910 will start to measure the fluorescein concentration in the sample.
1. Press the CAL on the main page, then choose Fluorescein and press the OK key to launch the fluorescein calibration page. 2. Follow the message prompts, insert the DI blank into the sample vial compartment and press the OK key. 3.
The middle range and high range calibrations from steps 4 to 8 are optional if only low range fluorescein measurement is intended. Figure 10 Low Range Fluorescein Calibration Figure 11 Middle Range Fluorescein Calibration Figure 12 High Range Fluorescein Calibration The standard solutions shall be stored in a brown or black opaque bottle.
5 Colorimetric Measurement 5.1 Supported Methods A wide range of colorimetric methods is supported by the SP-910 analyzer and the number of them keeps increasing with continuous development of Pyxis. See corresponding Hach® methods in Appendix A. Table 2 List of Supported Colorimetric Methods...
Page 21
Abbreviated Method Name Description Range Method Name OrgP Phosphonate UV digestion and ascorbic acid 7.1 ppm as reduction molybdenum blue PBTC method PAmi OPO4-Amino Reactive phosphate, amino 30.0 ppm acid reduction method ClO2 ClO2-DPD DPD method, USEPA accepted 5.00 ppm for reporting drinking water analysis ClO2D...
Figure 14 Single Timer Method 5.4 Single-Vial Procedure 1. Place the sample vial filled with the water sample in the SP-910 sample vial compartment and press the ZERO button. The SP-910 will display the page shown in Figure 14.
The following procedure is typical for two-vial methods: 1. Place the prepared blank into the SP-910 sample vial compartment and press the ZERO button to zero the instrument.
The SP-910 must be zeroed using the prepared blank before the last timing step. Thus, the last timer button will not be selectable until the SP-910 has been zeroed using the prepared blank. Multi-timing step Hach® methods require zeroing the colorimeter using the prepared blank after the last timing step is completed.
Page 26
Direct Reading Bleach Percent Method, 0 to 15% The SP-910 has a 365nm UV LED and other deep blue LEDs that can be used to directly measure bleach concentration in the range of 0 to 15%. No reagent is required for the method and the displayed result is the sodium hypochlorite concentration in percentage.
7. Gently shake the sample vial for 30 seconds. 8. Place the sample vial back to the sample compartment. Product concentration will be updated and plotted on the screen every 3 seconds. When the 3 minute timer is reached, the final concentration will be shown on the screen.
The following steps are used to calibration a method: 1. Use a calibration standard of known concentration. Follow the steps required by the method and note the value reported by the SP-910. 2. If the measured value differs from the known standard value, Press the CONFG button to launch the method configuration page.
2. Insert the sample vial to the sample vial compartment. 3. Slide the light shield cover to the closed position. 4. Press the Turb on the main page, then press the OK key, The SP-910 will start to measure the turbidity in the sample.
3. Slide the light shield cover to the closed position. 4. Press the CAL on the main page, then choose the Turbidity calibration and press the OK button to launch the Turbidity calibration page. (Figure 20) 5. Press the OK key to measure the deionized water 6.
Wavelength (nm) not used Note that the absorbance values measured with the SP-910 is generally smaller than those measured with a spectrophotometer equipped with a monochromatic light source or detector. The SP-910 absorbance values should, however, linearly correlate with the...
Thus, for any colorimetric system, The SP-910 absorbance follows Lambert-Beer law. 8 Update SP-910 firmware via Bluetooth Interface 8.1 Download and Install the Latest Nebula Software http://www.pyxis-lab.com/support.html#Software 1) Unzip the downloaded file and double click Nebula.exe to launch installation...
8.2 Turn on SP-910 Long press OK button for 5 seconds to turn off the SP-910 Press LEFT button, then click OK button while holding down LEFT button SP-910 displays below menu Figure 26 Bootloader Menu Click RIGHT button,SP-910 enters upgrading mode,SP-910 displays below menu to...
Nebula PC software if no device was found. 3) Once the device is connected, click the Load File button to select SP-910 firmware file, and click update button to start upgrading SP-910. It takes about 160 seconds to complete the procedure.
10 Calibrate a ST-500 with SP-910 The SP-910 can be used to verify the result of inline Pyxis ST-500 and other probes by measuring the sample water took from the inline probe sample line. The SP-910 can then be used to calibrate the inline probes over the Bluetooth connection.
Figure 29 Read Inline Device Use the SP-910 to measure the sample water by clicking Up key, as shown in Figure 30 Figure 30 SP-910 Measure Sample Water Click OK key to send the calibration instruction to the ST-500 via Bluetooth connection.
Figure 32 Inline Device Diagnosis Data In the cleanliness check page, please put the ST-500 probe into DI water and then click the OK button again to conduct cleanliness check. Figure 33 Cleanliness Check Figure 30 shows a probe may be fouled according to its diagnosis operational parameters.
Use a soft cloth or lint free paper tissue to clean the sample vial compartment periodically. Remove debris, scale, and deposit promptly. Although The SP-910 is protected from water damage, it is a good practice to avoid water entering the sample vial compartment and becoming trapping underneath the navigational control pad.
If an unspecific error occurs or the SP-910 cannot be turned on, reboot the instrument by taking a battery out of the battery holder and re-install the battery. If the SP-910 has been idle for more than two months and cannot be turned on, replace all four batteries with four new AA alkaline batteries.
Need help?
Do you have a question about the SP-910 and is the answer not in the manual?
Questions and answers