Lifting And Carrying The Transport Case; Checks To Perform Before Commissioning - Baur PD-TaD 62 User Manual

Portable diagnostics system
Table of Contents

Advertisement

Connecting the system
6.2

Lifting and carrying the transport case

 To transport the transport cases containing the PD-TaD 62 and accessories, pull them
along on their wheels.
To extend the telescopic handle of the transport case, release the locking mechanism (1)
and extend the telescopic handle (2).
 The transport cases and their contents weigh approx. 38.0 kg or 22.5 kg. Lifting or
carrying the transport cases requires two people.
6.3

Checks to perform before commissioning

1. Check all system components and mechanical connections for damage.
2. Check electrical connections and connection cables for damage.
Use only undamaged connection cables.
32 / 114
WARNING
Danger due to electric voltage, flashovers at the connection point,
or arcing fault on connection
Electric shock on touching live and active parts and due to residual
charges and induction voltages;
Burns, electro-ophthalmia, hearing damage.
 Use suitable personal protective equipment against electric shocks
and arcing faults.
 Observe the phase breaks.
 Ensure that adjacent live parts are secured against accidental
contact and flashovers with suitable covers (insulation mats,
insulating safety plates).
 You may touch the parts that were under voltage only if they are
visibly earthed and short-circuited.
PD-TaD 62
822-187-3

Advertisement

Table of Contents
loading

Table of Contents