Identifying problems

Please refer to the details of following symptoms or error messages. If the
problem persists, turn off the detector and consult your sales representative or
local dealer.
Symptom
The detector will
not turn on.
The status indi-
cator of the Sys-
tem Control Unit
does not light
up.
The status indi-
cator of the Sys-
tem Control Unit
does not light up
in green.
The green status
indicator lights
up and the or-
ange and blue
status indicators
are flashing.
WARNING:
Improper changes, additions, maintenance or repair of the
system can lead to personal injury, electrical shock and damage
to the equipment. Safety is only guaranteed when changes,
additions, maintenance or repairs are carried out by an Agfa
certified field service engineer. A non certified engineer
performing a modification or service intervention on a medical
device, acts on his own responsibility and makes the warranty
void.
Cause
The battery is not at-
tached.
The battery pack is not
charged.
The battery pack is bro-
ken.
The power cord is un-
plugged from the AC
outlet.
A hardware error has
occurred.
An error has occurred
during registration of
the DR Detector.
An error has occurred
during data communi-
cation.
DX-D 45C, DX-D 45G | Problem solving | 103
Remedy
Attach the battery.
Fully charge the battery pack.
Replace the battery pack.
Connect the plug to the AC out-
let firmly. If it still does not
work, replace the System Con-
trol Unit.
Turn off the System Control
Unit and turn it on again. If it
still does not work, replace the
System Control Unit.
Check the network connection
on the System Control Unit.
Check the network configura-
tion of the workstation.
Check if the System Control
Unit is turned on.
0292B EN 20190401 1627