Identifying problems

Please refer to the details of following symptoms or error messages. If the
problem persists, turn off the detector and consult your sales representative or
local dealer.
Symptom
The detector will not
turn on.
A fully charged bat-
tery is consumed
quickly.
The battery bay is
unusually hot.
Sharing the DR De-
tector between NX
Workstations fails
WARNING:
Improper changes, additions, maintenance or repair of the
system can lead to personal injury, electrical shock and damage
to the equipment. Safety is only guaranteed when changes,
additions, maintenance or repairs are carried out by an Agfa
certified field service engineer. A non certified engineer
performing a modification or service intervention on a medical
device, acts on his own responsibility and makes the warranty
void.
Cause
The battery is not at-
tached.
The battery pack is
not charged.
The battery pack is
broken.
The battery capacity
decreases.
The battery was
charged or used in
low temperatures.
The battery is mal-
functioning.
The DR Detector is
not configured on
the NX workstation.
DR 10s | Problem solving | 95
Remedy
Attach the battery.
Fully charge the battery pack.
Replace the battery pack.
The DR Detector battery can
deteriorate because of its char-
acteristics and structure. For
purchase of consumables, con-
tact your sales representative
or local dealer.
In low temperatures the bat-
tery capacity decreases. Use a
battery charged in normal
temperatures.
Stop using the battery and
consult your sales representa-
tive or local dealer.
Contact your local service rep-
resentative.
0351A EN 20190319 1349 *