Printer Output Test; Communication Port Test - GE DINAMAP ProCare 100 Service Manual

Hide thumbs Also See for DINAMAP ProCare 100:
Table of Contents

Advertisement

4.6.13 Printer Output Test

4.6.14 Communication Port Test

4-18
Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests
1. Load Thermal Paper into the print mechanism
2. Press PRINT button
3. Verify the printer outputs a record and print quality is good
1. Connect unit to a PC terminal emulator.
2. Turn unit on.
3. Type " NC0!E", press "Enter"
4. Verify response from unit " NC+!@"
5. Verify that within 40 seconds the pump starts.
6. Type " ND!5", press "Enter"
7. Verify response from unit " ND+!A"
8. Verify the pump stops.
ProCare Patient Monitor Service Manual
2009381-001
Revision A