Spo2 (Perform Only If Equipped With Spo2 Module) - GE DINAMAP ProCare 100 Service Manual

Hide thumbs Also See for DINAMAP ProCare 100:
Table of Contents

Advertisement

4.6.12 SpO
(Perform only if equipped with SpO
2
Revision A
Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests
9. Calibration verification is complete. Disconnect the probe simulator and
install the temperature probe. If the Monitor does not pass the calibration
verification, then the Monitor needs repair.
Note for Monitors equipped with Nellcor SpO
Nellcor is aware that various Nellcor oximetry platforms have some
interesting behaviors when used with various pulse simulators.
On occasion when testing the integrity of the Nellcor oximetry system,
abnormal results may occur when introducing large changes in the pulse
rate and/or pulse amplitude. Extreme changes in rate sent to the Nellcor
sensor by the SpO
miss in finding the pulse rate.
This is an expected result. To workaround this, incrementally step up or
down the settings on your SpO
and display the new pulse rate or saturation.
Nellcor recommends use of the SRCmax Portable Tester for use with
ProCare Monitors equipped with the Nellcor SpO
Masimo recommends BIO-TEK SpO
1. Connect the appropriate SpO
2. Verify the unit displays a:
Pulse value
Saturation value
Signal Strength Bar Graph
3. Disconnect the SpO
4. Verify the unit generates a
5. Press the SILENCE button.
6. Verify the sound has stopped but the error display remains
7. Re-connect the SpO
8. Verify the unit displays a:
Pulse Value
Saturation value
Signal Strength bar Graph
ProCare Patient Monitor Service Manual
simulator may cause the SpO
2
simulator and allow for the Monitor to detect
2
2
simulator and cable to the SpO
2
cable
2
alarm and speaker is sounding
sensor
2
2009381-001
module)
2
:
2
algorithm to completely
2
system.
2
simulators.
connector.
2
4-17