2.4. Host Communications Connector

2.4.1. DB15 Connector Pin Assignments

Revision A
Product Description: 2.4. Host Communications Connector
All host port signals are NON-ISOLATED and should be connected to
equipment conforming to IEC 601-1 ONLY. Where isolation of data
communication is required, the isolated level converter should be used. If
external alarm control is required, p/n 487208 (Isolated Remote Alarm Cable
Assembly) should ALWAYS be used. Please refer to the Information Sheet
included with the isolated remote alarm cable for operational details.
Note: When using remote alarm, the ProCare Monitor should be considered
the primary alarm source. The secondary alarm is used for secondary
purposes only.
Connection Details
Host Port Connector (rear panel)
WARNING! Auxiliary equipment
connected to the DINAMAP
Monitor will result in the formation of
an electromedical system and thus
must comply with the requirements of EN 60601-1-1/ IEC 601-1. All host port
signals are NON-ISOLATED and should be connected to equipment
conforming to IEC-601-1, configured to comply with IEC 601-1-1 ONLY.
Where isolation of data communication is required, GE Medical Systems
Information Technologies part number ILC1926 should be used. If external
alarm control is required, GE Medical Systems Information Technologies
part number 487208 (Isolated Remote Alarm Cable Assembly) should
ALWAYS be used. When a high-priority condition is displayed on the
Monitor, the remote alarm signal becomes active within 0.5 seconds. The
active state of the alarm signal is an open circuit. In the inactive state the
alarm signal is connected to ground. Please refer to the Information Sheet
included with the isolated remote alarm cable for operational details.
Note: When using remote alarm, the ProCare Monitor should be considered
the primary alarm source. The secondary alarm is used for secondary
purposes only.
Pin #
Function
1
Common
2
Inverted TTL Transmit Data
3
Inverted TTL Receive Data
4
+5 volts
5
No connection
6
No connection
7
Common
8
Remote Alarm
9
No connection
10
No connection
11
RS232 Transmit Data (TxD)
12
No connection
13
RS232 Receive Data (RxD)
14
No connection
15
No connection
ProCare Partient Monitor Service Manual
2009381-001
®
ProCare
2-9