Appendix A: Specifications
Thermal specifications for plates used in the instrument apply to flat-bottom plates with isolated
wells. All other plate specifications apply to standard 96-well polystyrene flat-bottom plates.
Note: Technical specifications are subject to change without notice. Molecular Devices
provides validation documentation for software and hardware, as well as absorbance
and fluorescence detection test tools with its SpectraTest® solutions. The SpectraTest
line of microplate reader validation packages provide automated and comprehensive
validation of a microplate reader's optical performance.
Absorbance Photometric Performance
Item
Wavelength Range
Wavelength Selection
Wavelength Bandwidth
Wavelength Accuracy
Wavelength Repeatability
Dynamic Range
Photometric Resolution
Photometric Accuracy Linearity
Photometric Precision
Stray Light
Light Source
Fluorescence Intensity Performance
Item
Filter Based
Dynamic Range
Sensitivity: Fluorescein
Excitation Wavelength and
Emission Wavelength Range
Adjustable Read Height
Read Time
5090339 A
Description
200-1000 nm
Monochromator, tunable in 1 nm increments
≤5.0 nm full width half maximum
2 nm across wavelength range
±0.2 nm
0 to 4 OD
0.001 OD
< ± 0.006 OD ± 1.0%, 0-3 OD
< ± 0.003 OD ± 1.0%, 0-3 OD
≤0.05% at 230 nm
Xenon flash lamp (5 Watts)
Description
Top Read
> 6 Decades
1 pM
For available Fluorescence filter cubes, see
Support on page
45.
Z range from upper plate @ 1 mm to max height 15.98 mm
30 seconds 96-well plate
96 seconds 384-well plate
A
Accessories and
41
Need help?