Preparation before use

☆ Note the following points to properly use the instrument.
Do not place the instrument in a
location exposed to direct sunlight,
or exposed to direct air-blow from
air-conditioning equipment, etc.
Display section
Do not push the
display section or the
operation-key section
with a pointed object.
Place the instrument on a solid floor.
If the installed place is not flat, adjust the levelness with the leveling leg to make it stable.
Keep the instrument away from any equipment that generates vibrations.
The rated operating temperature of this instrument ranges from -10°C to +40°C. Do not use the
instrument outside of this temperature range.
Unplug the power cable when the instrument is not in use or during inspection.
When you weigh products which are easy to charge, it is recommended that a ground earth should
be added (from a screw portion on the platform bottom cover).
Power cable (about 4 m
long) with a 100 VAC
plug
Ground wire (about 1 m long)
For safety, make sure to establish a
ground connection using the ground
wire attached to the power cable.
4
Firmly plug the power cable
into a receptacle of 100 V,
50/60 Hz power supply.
Spirit level
The instrument must
be leveled correctly.
Adjust its levelness
with the leveling leg
so that the air bubble
in the spirit level is
located inside the
red circle at the
center of the level.
Leveling leg