Section 3-Calibration
Introduction
This section provides calibration procedures for the Propaq Encore. These procedures allow you to
calibrate the Encore to Welch Allyn factory specifications. Calibrations procedures should be
performed only by qualified service personnel.
Calibration procedures must be performed whenever a question arises as to the performance of
the Encore, and as indicated in Section 1.
Functional verification procedures (see Section 2) verify that the monitor is performing to its
specifications. A functional verification (Section 2) must follow the calibration.
Note
Respiration, temperature, printer, and SpO
Sidestream CO
modules to support service of these options, perform a functional verification for
those associated modules.
Note
The mainstream CO
verification procedure of Section 2 AND you have eliminated all possible faults as
listed in the table .
Welch Allyn
has no user-performable calibration. After replacement of any
2
circuitry should be calibrated only if it fails the functional
2
Section 3
Calibration
options do not require calibration.