Accessories; Power Supplies; Cables - Teledyne HASTINGS Everywhereyoulook 200 Series Instruction Manual

Flowmeters/controllers
Table of Contents

Advertisement

201/203/205/207 Series Flowmeters and Controllers Instruction Manual

Accessories

1.5.1.

Power Supplies

Hastings power supplies are available in one or four channel versions. They convert 100, 115 or 230VAC
to the voltage required to operate the instruments and provide a digital indication of the flow rate.
Hastings power supplies are power limited to no more than 150 VA max (fused) and use a switch
compliant to EN 60601-1. Interface terminals for the analog output signals and a main power on/off
switch are located on the rear panel as well. For safety reasons it is recommended that the power
supply (and on/off switch) be kept near the flow instruments and that the on/off switch not be blocked.
Throughout this manual, when reference is made to a power supply, it is assumed the customer is using a
Hastings power supply. Designs and models offered as Hastings Power supplies change as new technology
becomes available. For knowledge of the most up-to-date Hastings power supplies, please consult either
Hastings sales staff or our website at
WARNING: To prevent the spread of fires in the event of a fault, only use power supplies power
limited to no more than 150 VA. The current available from a power supply used with a 200
series instrument should be no greater than 1.6 A. Hastings power supplies meet these
requirements.
1.5.2.

Cables

Cables to power Hastings flow instruments connect from the 15 pin "D" connector on the back of the
power supply directly to any of the 200 series flow instruments. More information about the available
cables can be found on our website at
EAR99 Technology Subject to Restrictions Contained on the Cover Page
http://www.teledyne-hi.com
http://www.teledyne-hi.com
Page 10 of 28