Table 2: Displayed values
Value
Pressure sensor at air outlet
Pressure
Flow sensor in flow generator*
Leak
Tidal volume
Respiratory rate
Minute ventilation
*
** The displayed values are estimates. They are provided for trending purposes only.
Table 3: Maximum low impedance flow at stated pressures
Pressure
(cm H
6.6
13.2
20.0
Note: The manufacturer reserves the right to change these specifications
without notice.
Symbols which appear on the product
36
Range
-5 to 30 cm H
0–120 L/min
100–3000 mL
6–60 BPM
0.6–60 L/min
Results may be inaccurate in the presence of leaks.
Flow
O)
(L/min)
2
177
202
202
Attention, consult accompanying documents
Class II equipment
Type CF equipment
Drip Proof
Start/Stop or Mask-Fit
Accuracy
O
±0.5 cm H
O
2
2
**
**
±0.5 BPM
**
Display Resolution
0.1 cm H
O
2
1 L/min
1 mL
0.1 BPM
0.25 L/min