Temperature Recorder; Usage Of Temperature Recorder; Setting Of Temperature Recorder - Phcbi MBR-107D Operating Instructions Manual

Blood bank refrigerator
Table of Contents

Advertisement

TEMPERATURE RECORDER

Usage of temperature recorder

The temperature recorder is operated for about one year with size "C" or "R14" battery.
recording paper can be used for one month. Temperature from +20
of 1
o
C. Since it is of a pressure sensitive type, it is not necessary to supply ink.
Temperature alarm setting
In window of this temperature recorder, upper limit alarm indicator (red needle) is set at about 6
lower limit alarm indicator (blue needle) is set at about 2
yellow needle (chamber temperature indicator) is between the other two needles, the chamber
temperature is normal. If the yellow needle is higher than the upper limit or lower than the lower limit,
the alarm device will be activated.

Setting of temperature recorder

As shown in the Fig. 1, pull the knob on the upper part of
the temperature recorder forward to change the recording
paper or battery.
Setting of recording paper
1. The information on the recording paper is shown in Fig.
2.
2. Pull the marking cartridge up after opening the top lid.
See Fig. 3. The lid can be opened by turning the knob
counterclockwise.
3. As shown in Fig. 4, insert the recording paper with the
"begin" tab placed in the marking cartridge. Check that
the printed side is facing out.
4. Place the recording paper beneath the arm and between
the plate spring and guide plate in the direction of the
arrow.
Note:
 Do not scratch or put pressure on the recording paper.
 Do no bend the recording paper.
 Do not reverse the recording paper manually.
o
C to -6
o
C. Both needles are fixed in place. When the
17
A roll of
o
C is recorded in increment
o
C and
Fig. 1
Fig. 2
Fig. 3
Fig. 4