Risk of contamination.
Potential contaminations will not remain in the centrifuge while the
device is operated.
WA RNIN G
Take appropriate protection measures to prevent spread of
contaminations.
A centrifuge is no closed containment.
Damage to health from centrifuging explosive or flammable
materials or substances.
Do not centrifuge explosive or flammable materials or substances.
WA RNIN G
Serious injuries can occur if you touch a spinning rotor with
your hands or tools.
A rotor can still be spinning after a power failure occurs.
WA RNIN G
Do not open the centrifuge before the rotor has stopped spinning. Do
not touch a spinning rotor. Open the centrifuge only if the rotor has
stopped spinning.
Never use your hands or tools to stop a spinning rotor.
The emergency lid release may be used in emergencies only to
recover the samples from the centrifuge, for example, during a power
failure
Risk of injuries from defective lid spring
Make sure that the centrifuge lid can be opened completely and that
it will stay in its position.
WA RNIN G
Regularly check the lid springs for their proper function.
Do not operate the centrifuge with a defective lid spring.
Let an authorized service technician replace defective lid springs.
("Mechanical Emergency Lid Release" on page
9 / 94
67).