Error
Description
E-24
The centrifuge cannot
be operated.
E-29
Motor does not start
E-31
Rotor stops without
deceleration to standstill
or does not start.
E-33
Rotor stops with
deceleration to standstill.
E-36
Rotor stops with
deceleration to standstill.
The centrifuge cannot
be operated.
E-41 - E-56
Rotor stops with
deceleration to standstill.
The centrifuge cannot
be operated.
Troubleshooting
Solutions
Wrong status information from the lid latch.
• Switch the instrument off and on again.
• After re-switching on, the display shows Lid FAiL.
• If the centrifuge lid has been already opened, the display
shows CLOSE Lid. Close the lid.
• The centrifuge tries to open the lid to switch for starting the
normal operation mode.
If the error message still shows, contact a service technician.
Motor or rotor blocked
• Switch instrument off and on again using the mains switch.
• Open the centrifuge lid.
• Check whether the rotor can turn freely.
If the error message still shows, contact a service technician.
Overtemperature in the motor
• Turn instrument off and unplug mains power cord.
• Check and clean the venting slots if necessary and
respectively the filter unit of the cooled centrifuge.
• After approx. 60 minutes you can restart the instrument.
Observe the maximum permissible environmental temperature.
If the error message still shows, contact a service technician.
Overpressure in the refrigeration system
• Turn instrument off and unplug mains power cord.
• Check and clean the venting slots if necessary and
respectively the filter unit of the cooled centrifuge.
• After approx. 60 minutes you can restart the instrument.
Observe the maximum permissible environmental temperature.
If the error message still shows, contact a service technician.
Overcurrent or error in current measurement
Switch the instrument off and on again.
If the error message still shows, contact a service technician.
Internal program error
Switch the instrument off and on again.
If the error message still shows, contact a service technician.