Fluke 1521 User Manual page 44

Handheld thermometer readout
Table of Contents

Advertisement

1521 Handheld Thermometer Readout
User's Guide
Description
Set probe type
Read ITS-90 R(.01) or CVD R0 r[0]
Set ITS-90 R(.01) or CVD R0
Read ITS-90 coefficient
Set ITS-90 coefficient
Read CVD coefficient
Set CVD coefficient
Read THERM thermistor
coefficient
Set THERM thermistor
coefficient
Read probe wires
Set probe wires
Read probe calibration date
Set probe calibration date
Read probe due date
Set probe due date
Read filter
Set filter
Read low range calibration
Set low range calibration
Read high range calibration
Set high range calibration
Read meter calibration date
Set meter calibration date
Read meter due date
Set meter due date
Legend:
40
Command
Example
p[robe]=IT[S-90]/
pr
IE[C751]/ C[VD]/
Y[SI400]/ T[HR]
r0
r[0]=n
r0=100.106
a/b/c/d/a4/b4
a
a/b/c/d/a4/b4=n
a=-2.8644E-5
al[pha]/de[lta]/be[ta]
al
al[pha]/de[lta]/be[ta]=n al=0.0038512
b0/b1/b2/b3
b3
b0/b1/b2/b3=n
b3=-9.944E-12
wi[res]
wi
wi[res]=2/3/4
wi=4
pc[al]
pcal
pc[al]=yyyy-mm-dd
pc=1999-03-28
pd[ue]
pdue
pd[ue]=yyyy-mm-dd
pd=1999-03-28
fi[lter]
fi
fi[lter]=n
fi=0
*c1
*c1
*c1=n
*c1=-0.012
*c2
*c2
*c2=n
*c2=-0.9
mc[al]
mcal
mc[al]=yyyy-mm-dd
mc=1999-03-28
md[ue]
mdue
md[ue]=yyyy-mm-dd
md=1999-03-28
[] Optional Command data
{} Returns either information
Response
Response
Format
Example
r0: 9999.999
r0: 100.035
{a/b/c/d/a4/b4}:
a: -2.8644101E-5
-9.999999E-99
{al/de/be}: 9.99999999 al: 0.0038123
{b0/b1/b2/b3}:
b3: -9.944070e-12
9.999999e-99
wi: 9
wi: 4
pc: yyyy-mm-dd
pc: 1999-03-28
pd: yyyy-mm-dd
pd: 1999-03-28
fi: 99.99999
fi: 0.5
c1: 99.999
c1: -0.0121
c2: 9999.9
c2: -0.9
mc: yyyy-mm-dd
mc: 1999-03-28
md: yyyy-mm-dd
: 1999-03-28
Range
ITS-90,
IEC751,
CVD, YSI400,
THERM
1 to 900
1.0 to 9.9e-18
E+/-18
E+/-18
2,3,4
1999-01-01 to
2100-12-31
1999-01-01 to
2100-12-31
0.0 to 60.0
-1.999 to 1.999
-1980.0 to
1980.0
1999-01-01 to
2100-12-31
1999-01-01 to
2100-12-31