6.2.2 Performing QC

QC provides 2 reference measurement tests Level 0 and defined NO with QC tester.
In the "6.4.1 Quality Control (QC) setting", QC can be restricted to 1 QC reference measurement only. Dependant on the QC
settings, QC is performed first with Level 0 and then with defined NO with QC tester, or with the selected reference QC-
measurement only.
Start QC‑measurement
QC Tester
Language
Jeppe M. Gregersendan
Administrator
Konsta Haapakoski
Professional
Oona Pakarinen-Kattilakoski
Professional
3. Perform a regular measurement.
Q U A L I F Y
After the measurement the result can be:
Qualify: Status during first 3 measurements of QC tester qualification.
Passed: Result is within acceptance range of the mean of the last
3 measurements.
Passed
Failed: Result is outside acceptance range of the mean of the last
Passed
3 measurements.
When the qualification procedure is passed, the qualification status of the QC
tester is changed to Qualified.
The qualified QC tester can perform the defined NO reference measurement test.
4. In the Settings menu, select PERFORM QC.
The QC Tester screen opens (only if QC tester is activated).
In the list, you see all users with their QC status. Your account is highlighted
Qualified
and can be selected.
Qualified
If no user management is enabled only the administrator is shown in the list.