13. Trigger Settings (Capturing Target Data to be Recorded)

13.2. Manual Trigger/External Trigger

Regardless of trigger modes, trigger is activated manually or externally, thereby initiating the recording.

13.2.1. Manual Trigger

Pressing the Manual/Marking button on the operation panel generates a trigger irrespective of other
trigger settings.

13.2.2. External Trigger (TRIG IN)

Trigger is generated by the falling edge of a 0 to 5-V signal. To use, enter a signal into TRIG IN in
the recorder.

13.2.3. External Trigger Output (TRIG OUT)

When a trigger is generated, the L level of the signal, which has TTL level and 10-ms pulse width, is
output from the TRIG OUT terminal.
Press Manual Trigger/Marking. A trigger is generated and
trigger LED (green) comes on.
TRIG OUT
TRIG IN