To get into the menu press the
tool button. Select the required
point by using the „+" or „-„
button. Press the „OK" button.
To get into the menu press the
tool button. Select the required
point by using the „+" or „-„
button. Press the „OK" button.
To get into the menu press the
tool button. Select the required
point by using the „+" or „-„
button. Press the „OK" button.
NB: All changed values have to be confirmed by pressing the " OK " button.
Otherwise the existing values are valid.

11.0 OPERATING INSTRUCTIONS

a) Switch on the power supply.
b) Allow a period of time for water entering the sample tube.
c) Flow oil free water through the system for a few minutes and check that the
display show 0 to 2 ppm. If not, clean proper before adjusting the unit
according section 10.4 "Service - Offset".
d) Switch the instrument sample supply from the clean water supply to the
separator sampling point connection.
e) The instrument is now ready for use.
Issue: 27.05.08
DECKMA HAMBURG GmbH
The temperature of the
measuring cell and the sample
water will be shown
The details of the measuring
cell will be shown.
Information about the software
version and the web address
will be shown.